MA Performance Test
The performance test can be used to check whether the analyzer has a fault
or is ready for operation. During a performance test, both the heating unit and
the weighing system are tested at the same time. You will need a ReproEasy
pad (accessory) in order to carry out a performance test.
Starting the Performance Test
Calibration/Adjustment
MA performance test
Weighing system adjustment
Calibration report
t Select “MA performance test.”
y The display switches to a view with three heating temperatures.
Please ensure that the device has not been heated for at least one hour
before a performance test is carried out.
MA performance test
The device is too warm for a performance
test.
The device is still cooling down.
t Select “m” to confirm the message and start the performance test again later.
MA performance test
Temperature 70 °C
Temperature 100 °C
Temperature 130 °C
t Select one of the suggested temperatures given for the method you are using.
Open hood
0.000 g