38
Moisture Analysis
Operating Instructions Moisture Analyzer MA37
Printing Out the Analysis Report
If a lab printer is connected, you can print out the result of the moisture analysis.
a
t Select a.
y The analysis report is printed out.
The settings for the analysis report can be configured in the system settings
(see section “Printout” in Chapter “System Settings”).
Further information and an example of an analysis report can be found in
Chapter “ISO/GLP-compliant Printout”.
Removing a Sample
Danger of burns due to hot sample pan!
The sample and sample pan may be very hot.
Do not touch the sample pan.
Use sample tongs (accessory) to remove the sample.
Remove sample
Remove old sample from sample
chamber.
y When you have exited the measurement by selecting OK, you can remove the
sample from the moisture analyzer.
t Open the hood.
t Remove the sample pan from the moisture analyzer using sample tongs.
t Close the hood.
t Allow the sample to cool down and dispose of it properly.