Care and Maintenance
Service
To ensure its continued weighing accuracy, we recommend performing regular 
servicing on your analyzer at least once a year. The Sartorius Service Center offers 
different service contracts for this purpose, which can be individually adapted to 
suit your needs (see also www.sartorius.com/service). Each service produces a 
calibration certificate. The analyzer and its connections must undergo a technical 
safety inspection on a regular basis carried out by an electrician (e.g., every 2 years).
Repairs 
Repair work must only be carried out by trained service technicians. The analyzer 
must be unplugged during repair work. Unplug the power cable from the socket. 
Failure to do so may affect the measurement accuracy of your analyzer and pose 
aserious risk to the user. Contact the Sartorius Service Center or a Sartorius dealer 
to arrange for proper repairs.
Cleaning the Analyzer
Cleaning the Control Panel
Switch the display to standby mode to avoid modifying the settings for operation 
during cleaning.
t  Select the menu key to switch to the method menu. 
t  If you then select j, the display is switched off. 
t  To switch the display back on: Select j at the bottom left of the display. 
Themoisture analyzer starts in the application most recently used before the 
analyzer was last switched off.
Cleaning the Housing
Disconnect the analyzer from the power before cleaning. Disconnect any data 
cables from the moisture analyzer. 
 
Never open the analyzer housing. The parts contained in the housing cannot be 
cleaned, repaired, or replaced by the operator. 
–  Make sure that no liquid or dust gets into the analyzer.
–  Remove the disposable sample pan and pan support from the analyzer.
–   Never use cleaning agents that contain solvents or abrasive ingredients, which 
can ultimately damage the equipment. 
t   Clean the outside of the housing with a lint-free cloth and mild cleaning agent 
(e.g., isopropanol). 
t  Then dry the analyzer with a soft cloth.