EasyManua.ls Logo

Sartorius MA37 - Cleaning the Heating Module

Sartorius MA37
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
58
Care and Maintenance
Operating Instructions Moisture Analyzer MA37
Contaminated Equipment:
Health risk from product contamination due to product deposits and the
collection of residue with microbial contamination.
Health risk from biological or microbiological substances.
Observe cleaning specifications.
Check the cleaning result closely.
Cleaning the Heating Module
Danger of Burns:
The inner part of the heating module and parts of the sample chamber may be
extremely hot.
Wait until the heating module has cooled completely.
Avoid touching the heating element. To remove spray residues or deposits from
the heating element, use a weak solvent such as ethanol.
Unlocking the Heating Module
1) Open the heating module all the way.
2) Pull the unlocking handle at the back of the heating element to unlock the
heating element.
Removing the Heating Module
3) Pull the heating module up out of the guide.

Related product manuals