4) Clean the heating element and temperature sensor with a weak solvent such as
ethanol.
5) If necessary, pull the air extractor grate out of the heating module. Then clean
the heating module and grate in a dishwasher.
6) Once cleaning is complete, reassemble the heating module and reinstall in the
analyzer, following the instructions in reverse order. The heating module must
snap into place on the analyzer.
Cleaning the Bottom of the Sample Chamber
Danger of Burns:
Parts of the sample chamber may be extremely hot.
Wait until the bottom of the sample chamber has cooled completely.
Removing the Sample Chamber Bottom
1) Unlock the “Locking socket (bayonet)” as in general view with a suitable coin.