Warning  and 
Safety Instructions
  Signum scales comply with the 
European Council Directives as well 
as international regulations and 
standards for electrical equipment, 
electromagnetic compatibility, and the 
stipulated safety requirements. 
§ Read these operating instructions 
thoroughly before using your device. 
That way you will prevent damage to 
the equipment.
! Do not use this equipment in hazardous 
areas.
! The indicator may only be opened by 
authorized service technicians who 
have been trained by Sartorius and who 
follow Sartorius’ standard operating 
procedures for maintenance and repair 
work.
! Make absolutely sure to unplug the 
indicator from power before you 
connect or disconnect any electronic 
peripheral devices to or from the 
interface port.
! If you use electrical equipment in 
installations and under ambient 
conditions requiring higher safety 
standards, you must comply with the 
provisions as specified in the applicable 
regulations for installation in your 
country.
! The display value can be affected by 
extreme electromagnetic influences. 
Once the disturbance has ceased, 
the instrument can be used again in 
accordance with its intended purpose.
–  Warning when using pre-wired RS-232 
connecting cables: RS-232 cables 
purchased from other manufacturers 
often have pin assignments that are 
incompatible with Sartorius products. 
Be sure to check the pin assignments 
against the chart in this manual before 
connecting the cable, and disconnect 
any lines identified differently from 
those specified by Sartorius.
–  Use only extension cords that meet 
the applicable standards and have a 
protective grounding conductor.
–  Disconnecting the ground conductor 
is prohibited.
–  Note on Installation:  
The operator shall be responsible for 
any modifications to the equipment 
and for any connections of cables or 
equipment not supplied by Sartorius 
and must check and, if necessary, 
correct these modifications and 
connections.  
Information on operational quality is 
available on request from Sartorius 
(in line with norms pertaining to 
immunity).
$ If there is visible damage to the 
equipment or power cord: unplug the 
equipment and lock it in a secure place 
to ensure that it cannot be used for the 
time being.
  IP Rating
–  All models are rated to IP43
–  Only connect Sartorius accessories and 
options, as these are optimally designed 
for use with your Signum scale.
$ Do not unnecessarily expose the device 
to aggressive chemical vapors or to 
extreme temperatures, moisture, shocks, 
or vibration.
–  Only clean the device as stipulated in 
the cleaning instructions: see “Care and 
Maintenance."
$ If you have any problems with your 
Signum scale:  
contact your local Sartorius office, 
dealer or service center.
  IP Protection: 
–  The protection rating for the display 
and control unit is ensured only if 
the rubber gasket is installed and all 
connections are fastened securely 
(including the caps on unused sockets).  
Weighing platforms and equipment 
must be installed and tested by a 
certified technician.
–  If you install an interface port or 
battery connector after setting up 
your Signum, keep the protective 
cap(s) in a safe place to be used for 
protecting the interface port or battery 
connector when not in use, or prior to 
shipment. Do not leave the interface 
port uncovered. If you are not using a 
particular connector, replace the cap to 
protect the data interface from vapors, 
moisture and dust or dirt.
  Using the Equipment in Legal 
Metrology:
–  If the scale is to be verified, make sure 
to observe the applicable regulations 
regarding verification. 
–  If any of the verification seals are 
damaged, make sure to observe the 
national regulations and standards 
applicable in your country in such 
cases. In some countries, the equipment 
must be re-verified.