91
GMP-compliant
Printouts
When the corresponding menu item is
active, the printout is bracketed by a
GMP header and a GMP footer (GMP:
“Good Manufacturing Practice”).
The GMP header precedes the first
measured result. The GMP footer is
printed after the last result in a series
of measurements (“GMP-compliant
printout always for several application
results,” menu item 7.13.3).
To end a series of measured results,
press and hold the p key. In this case,
the A symbol
is displayed after the GMP header is
printed and remains in the display until
the GMP footer is printed.
If you toggle to a different platform
(Signum
®
2 and 3 only) while a GMP
printout of several measured results is
being generated (menu item 7.13.3),
the GMP footer for the platform used
up to that point is generated when you
press the n key. The GMP header for
the other platform is included on the
next printout generated.
A GMP-compliant printout is generated
automatically at the conclusion of
calibration/adjustment, linearization
routines, as well as when you set or
clear a preload.
Three examples of GMP headers and one example of a footer are shown in the following.
On Signum
®
1 models, the “date and time” line is not included.
Weighing platform WP 1:
------------------- Dash line
14.01.2011 09:43 Date/Time
1
)
Type SIWR Signum
®
Type
Ser.no. 12345678 Signum
®
Serial no.
Vers. 1.1007.12.1 Software version of application
BVers. 01-25-01 Software version of basic version
------------------- Dash line
Weighing platform WP 2 (xBPI protocol):
2
)
------------------- Dash line
14.01.2011 09:45 Date/Time
1
)
Type SIWR Signum
®
Type
Ser.no. 12345678 Signum
®
serial no.
Vers. 1.1007.12.1 Software version of application
BVers. 01-25-01 Software version of basic version
Type IS12000S Platform type
Ser.No 12345678 Platform serial no.
------------------- Dash line
Weighing platform WP 2 (SBI protocol):
2
)
------------------- Dash line
14.01.2011 09:45 Date/Time
1
)
Type SIWR Signum
®
Type
Ser.no. 12345678 Signum
®
serial no.
Vers. 1.1007.12.1 Software version of application
BVers. 01-25-01 Software version of basic version
Type SBI (Platform type)
------------------- Dash line
GMP footer:
------------------- Dash line
14.01.2011 09:45 Date/Time
1
)
Name: Field for signature
Space line
------------------- Dash line
1
) Only if clock module is available
2
) Only for Signum
®
2 and 3 display and control units