Net-total Formulation Application R
75
  Net-total Formulation  
 Application R
  With this application, you can weigh in 
different components up to a defined 
total. Each component is saved in the 
net-total memory.
 Characteristics
–  Weigh in up to 999 components in 
series
–  Net-total formulation cannot be 
combined with a level 1 or level 2 
application
–  Current component number displayed 
in the text lines (indicating the 
component to be added)
–  Toggle the display from “component 
mode” to “additive mode” by pressing 
the w key
  –  Component mode: Display the weight 
of the component currently on the 
platform (for 1 second after it is 
saved; then the platform is tared)
  –  Additive mode: Display the weight of 
all components on the platform (after 
it is saved, the net weight of the last 
component added is displayed briefly)
–  Toggle to a second weighing platform 
while weighing on the first
–  Activate Info mode via the I key
–  Automatic component printout when it 
is saved. Configured in Setup under:
Appl: R
: 3.17.
Signum
®
 3:
  appl 3: 
Net Tot: 3.17.
  If the 3.17.2 menu item is selected, the 
entire component record is printed. 
If the 3.17.3 menu item is selected, 
the following items are generated only 
once for the first component: 
blank line, date, time, ID1 through ID4, 
header lines 1 and 2. For subsequent 
components, each “component” item 
(“Comp xx”) is followed by a blank line.
–  Automatic taring of container weight. 
Configured in Setup under:
Appl: R
: 3.7.
Signum
®
 3:
  Appl: a.tare: 3.7.
–  Restore factory default settings. 
Configured in Setup under:
Appl: R
: 9.1.
Signum
®
 3:
  Appl: def.app: 9.1.