Net-total Formulation Application R
76
 Preparation
§ Switch on the scale: Press the e key
§ While all segments are lit, press the 
) key
§ Select the Setup menu: Press the 
k key repeatedly until appl is 
displayed
§ Confirm appl: Press the ) key
§ Selecting the Net-total Formulation 
Application:
  Press the k key repeatedly and 
confirm with the ) key
  Net-total Formulation Application 
Parameters
    3.6.  Minimum load for autosave
      3.6.1*  1 digit
      3.6.2  2 digits
      3.6.3  5 digits
      3.6.4  10 digits
      3.6.5  20 digits
      3.6.6  50 digits
      3.6.7  100 digits 
    3.6.8  200 digits
      3.6.9  500 digits
      3.6.10  1000 digits
    3.17.    Individual/Component printout 
when saved
      3.17.1   Automatic printing off
      3.17.2*   Print the entire standard 
print configuration 
every time with the O 
key
      3.17.3   Print the entire standard 
print configuration 
once with the O key
  * = Factory setting
§ Save the settings with the ) key and 
exit Setup: Press the ( key several 
times.
  Minimum load
  The minimum amount that a 
component must weigh before it can be 
saved in net-total memory is configured 
in Setup under:  
Appl: R
: 3.6.
Signum
®
 3:
appl 3
  
Net Tot: 3.6.
  Once the limit is exceeded by the load, 
the value can be saved. If the load on 
platform is too light, the following will 
occur when you try to save a value:
–  The error code Inf 29 is displayed
–  The weight is not saved
  The minimum load required for 
automatic taring of the container 
weight on the platform (first weight) 
is configured in Setup under: 
 
Appl: R: 3.5.
Signum
®
 3:
Appl: m.weigh: 3.5.
  You can choose from the following 
10 levels for this setting:
  1 digit
  2 digits
  5 digits
  10 digits
  20 digits
  50 digits
  100 digits
  200 digits
  500 digits
  1000 digits
  The “digits” here refer to the scale 
intervals in the connected weighing 
platform. If the interval of the 
connected platform is 1 g, for example, 
and 1000 digits are required, you must 
place at least 1000 g (=  1000 intervals 
=1000 digits) on the weighing platform 
for initialization.
  Net-total Formulation 
with Two Weighing Platforms
  This mode is used for weighing large 
and small components at the same 
time.
  It is possible to toggle from the small-
component platform to the large-
component platform once during a 
measurement series. Once you toggle 
to the large-component platform, the 
( and ) keys are available until 
a component is saved. For example, 
you can tare a partially-filled container 
taken from the small-component 
platform on the large component 
platform.
  The value in component memory 
on the small-component platform is 
transferred to the large-component 
platform and the weight unit is 
converted, if necessary. The Component 
and Additive display modes are both 
available on the large-component 
platform.
  The value read by the active platform 
is saved in component memory. 
The displayed result is accurately 
calculated in the active weight unit.
 
  When you press c to stop a 
measurement series the tare memories 
for both platforms are cleared, unless 
the large-component platform is an SBI 
instrument, in which case the platform 
is only tared.