77
Example:
Weighing in 3 components of a formulation recipe.
Settings (different from the factory settings):
Setup: Application 3: Net-total Formulation
Setup: Printout, PRTPROT: 7.7. Printer 1:
Select “Component log: Auto printout”, then select the menu line items of your choice
Setup: Printout, PRTPROT: 7.8 Printer 1:
Select “Total data record: Printout as per CF key,” then select the menu line items of your choice
1.) Place empty container on the platform
)
2.) Tare the scale
Note: If the automatic tare function is enabled, you do not need to press the ) key to tare the
platform; the tare weight is saved automatically when you place the container on the platform
A prompt to fill and save the first component is shown
3.) Add the first component to the container (in this example, 1100 g)
The weight of the first component is displayed
O
4.) Save the first component weight
Cmp001+ 1.100 kg The component record is printed automatically