45
 Preparation
§ Switch on the scale: Press the e key
§ While all segments are lit, press the 
) key
§ Select the Setup menu: Press the 
k key repeatedly until Setup is 
displayed
§ Open the Setup menu:  
Press the ) key
$ Select weighing platform 1 “WP1:"  
Press the ) key or
$ Select interface 1 “COM-1" or interface 
2 “unicom" (depending on the 
interface): Press the k key
  Select weighing platform 2 “WP2:"  
Press the ) key
    
Setup
      wp-1 
1
        1.9.       Calibration and Adjustment
          1.9.1 *    Ext. calibration/adjustment; default weight
          1.9.3    External calibration/adjustment; weight can be selected under menu item 1.18.1
          1.9.4    Internal calibration/adjustment (models with built-in motorized calibration weight only)
          1.9.8    Setting the preload
          1.9.9    Deleting the preload
          1.9.10    No function when you press the J key
        1.10.       Calibration/Adjustment sequence
          1.10.1    Calibration with automatic adjustment
          1.10.2 *    Calibration with manual adjustment
        1.11.       Zeroing range
          1.11.1    1 percent/max. cap.
          1.11.2 *    2 percent/max. cap.
        1.12.       Initial zero point range
          1.12.1    Factory setting (depending on the model type)
          1.12.2    2 percent/max. cap.
          1.12.3 *    5 percent/max. cap.
        1.13.       Tare/zero at power on:
          1.13.1 *    On
          1.13.2    Off, load previous tare values
          1.13.3    Only zero at power on
        1.15.       Adjustment prompt
          1.15.1 *    Off
          1.15.2    Calibration prompt W flashes on the display
        1.16.       External calibration 
1
)
          1.16.1 *    Activated
          1.16.2 
2
)    Blocked
        1.17.      Calibration weight unit
          1.17.1    Grams
          1.17.2 *    Kilogram
          1.17.3    Pounds 
1
)
        1.18.       Enter calibration weight
          1.18.1    External user calibration weight (enter value, e.g.: 10,000g)
§ Save settings with the ) key and exit Setup 
  by pressing the ( key repeatedly
)
1
) = Not available on scales verified for use in legal metrology
2
) = Factory setting for use in legal metrology
*  Factory settings