45
Preparation
§ Switch on the scale: Press the e key
§ While all segments are lit, press the
) key
§ Select the Setup menu: Press the
k key repeatedly until Setup is
displayed
§ Open the Setup menu:
Press the ) key
$ Select weighing platform 1 “WP1:"
Press the ) key or
$ Select interface 1 “COM-1" or interface
2 “unicom" (depending on the
interface): Press the k key
Select weighing platform 2 “WP2:"
Press the ) key
Setup
wp-1
1
1.9. Calibration and Adjustment
1.9.1 * Ext. calibration/adjustment; default weight
1.9.3 External calibration/adjustment; weight can be selected under menu item 1.18.1
1.9.4 Internal calibration/adjustment (models with built-in motorized calibration weight only)
1.9.8 Setting the preload
1.9.9 Deleting the preload
1.9.10 No function when you press the J key
1.10. Calibration/Adjustment sequence
1.10.1 Calibration with automatic adjustment
1.10.2 * Calibration with manual adjustment
1.11. Zeroing range
1.11.1 1 percent/max. cap.
1.11.2 * 2 percent/max. cap.
1.12. Initial zero point range
1.12.1 Factory setting (depending on the model type)
1.12.2 2 percent/max. cap.
1.12.3 * 5 percent/max. cap.
1.13. Tare/zero at power on:
1.13.1 * On
1.13.2 Off, load previous tare values
1.13.3 Only zero at power on
1.15. Adjustment prompt
1.15.1 * Off
1.15.2 Calibration prompt W flashes on the display
1.16. External calibration
1
)
1.16.1 * Activated
1.16.2
2
) Blocked
1.17. Calibration weight unit
1.17.1 Grams
1.17.2 * Kilogram
1.17.3 Pounds
1
)
1.18. Enter calibration weight
1.18.1 External user calibration weight (enter value, e.g.: 10,000g)
§ Save settings with the ) key and exit Setup
by pressing the ( key repeatedly
)
1
) = Not available on scales verified for use in legal metrology
2
) = Factory setting for use in legal metrology
* Factory settings