Z
53
Preparation
§ Switch on the scale: Press the e key
§ While all segments are lit, press the
) key
§ Select the Setup menu: Press the
k key repeatedly until appl is
displayed
§ Confirm appl: Press the ) key
§ Select the Counting application:
Press the k key repeatedly and
confirm with the ) key
Counting application parameters
3.6. Minimum load for initialization
3.6.1* 1 digit
3.6.2 2 digits
3.6.3 5 digits
3.6.4 10 digits
3.6.5 20 digits
3.6.6 50 digits
3.6.7 100 digits
3.6.8 200 digits
3.6.9 500 digits
3.6.10 1000 digits
3.9. Resolution for calculation
of reference value
3.9.1* Display accuracy
3.9.2 Display accuracy +1
decimal place
3.9.3 Display accuracy +2
decimal places
3.9.4 Internal resolution
3.11 Parameter for saving weight values
3.11. 1* With stability
3.11. 2 With increased
stability
3.12. Average piece weight updating
3.12.1 Off
3.12.3* Automatic
3.13. Reference weighing instrument
3.13.1* No weighing platform
selected
3.13. 2 Weighing platform
WP 1
3.13. 3 Weighing platform
WP 2
* = Factory setting
§ Save the settings with the ) key and
exit Setup: Press the ( key several
times.
Parameter for saving weight values
The weight on the platform is saved as
a reference value when the platform
has stabilized. “Stability” is defined as
the point at which the fluctuation of
a measured value lies within a defined
tolerance range. The narrower the
tolerance range, the more stable the
platform is at “stability”.
In Setup, under:
Appl:
Z: 3.11.
Signum
®
3:
appl1: count: 3.11.
You can define whether the value is
saved when “standard stability” is
reached, or only at “increased stability”
(narrower tolerance range). If you
select “increased stability,” the value
saved for average piece weight will be
more accurate and the results more
reproducible, but the response time of
the weighing platform might be longer.
Accuracy of Average Piece Weight
Calculation
The resolution applied for calculating
the reference weight is defined in Setup
under:
Appl:
Z: 3.9.
Signum
®
3:
Appl1
: count: 3.9.
The resolution for calculating the
reference weight is increased if
“+1 decimal place,” “+2 decimal
places” or “Internal resolution” is
selected. With the “+1 decimal place”
setting, the net value is determined
to one additional decimal place (i.e.,
display accuracy x 10); “+2 decimal
places” increases display accuracy +
100, and “Internal resolution” uses
the maximum resolution available.
Minimum load
The minimum load required for
initialization of the weighing platform
is configured in Setup under:
Appl:
Z: 3.6.
Signum
®
3:
appl1:
count: 3.6.
Once the limit is exceeded by the load,
initialization can begin.
If the load on platform is too light, the
following will occur when you try to
save a value:
– The error code Inf 29 is displayed
– The weighing platform is not initialized
– The preset reference sample quantity is
saved
The minimum load required for
automatic taring of the container
weight on the platform (first weight)
is configured in Setup under:
Appl:
Z: 3.5.
Signum
®
3:
appl
: m.weigh: 3.5.
You can choose from the following
10 levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform. If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (= 1000 intervals
=1000 digits) on the weighing platform
for initialization.