Signum 1-2-3
  Classification Application W  Classification Application 
W
70
  Minimum load 
  The minimum load for the first class is 
configured in Setup under:
Appl: 
W: 3.6.
Appl 2
:  
Signum
®
 3:
class
: 3.6.
  Once the limit is exceeded by the load, 
initialization can begin.
  Once the application is initialized, 
a weight value below the minimum 
load is designated Class   0; no class is 
displayed.
  The minimum load required for 
automatic taring of the container 
weight on the platform (first weight), 
or for automatic printout of results, 
is configured in Setup under:
Appl: 
W: 3.5.
Signum
®
 3:
  Appl: m.weigh: 3.5. 
  You can choose from the following 
10 levels for this setting:
  1 digit
  2 digits
  5 digits
  10 digits
  20 digits
  50 digits
  100 digits
  200 digits
  500 digits
  1000 digits
  The “digits” here refer to the scale 
intervals in the connected weighing 
platform. If the interval of the 
connected platform is 1 g, for example, 
and 1000 digits are required, you must 
place at least 1000 g (=1000 intervals 
=1000 digits) on the weighing platform 
for the first class to activate autotaring 
or autoprint.
 Display
  The result of a given measurement is 
shown as either a weight value or a 
class number.
–  Weight display:
  The current weight is shown in the 
measured value line and the current 
class in the text lines.
–  Display of classes:
  The current class is shown in the 
measured value line, and the current 
weight in the text lines.
  Digital I/O  
Interface
  The Classification application supports 
the digital input/output-interface.
  There are 4 control lines, or outputs, 
which are activated as follows (see also 
the diagram below):
–  With 3 classes:
  – Class 1
  – Class 2
  – Class 3
  – Set
–  With 5 classes:
  – Classes 1/2
  – Classes 2/3/4
  – Classes 4/5
  – Set
  In Setup, under:
Appl: 
W: 4.7.
Signum
®
 3:
  Appl 2: 
class: 4.7.
  you can define whether these control 
ports are
–  switched off 
–  always on 
–  on at stability 
  The “SET” output normally changes 
its voltage level when the current 
weight exceeds the minimum load. 
Alternatively, you can assign the 
“Ready for use” function to this port. 
Configured in Setup under:
Appl: 
W: 4.3.
Signum
®
 3:
  Appl 2: 
class: 4.3.
70
Minimum Load 
The minimum load for the first class is
configured in Setup under:
Appl:
W: 3.6.
Appl 2
:
Signum
®
3:
class
: 3.6.
Once the limit is exceeded by the load,
initialization can begin.
Once the application is initialized, a
weight value below the minimum load
is designated Class  0; no class is
displayed.
The minimum load required for
automatic taring of the container
weight on the platform (first weight),
or for automatic printout of results,
is configured in Setup under:
Appl:
W: 3.5.
Signum
®
3:
Appl
: m.weigh: 3.5.
You can choose from the following
10 levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits" here refer to the scale
intervals in the connected weighing
platform. If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (=1000 intervals
=1000 digits) on the weighing platform
for the first class to activate autotaring
or autoprint.
Display
The result of a given measurement is
shown as either a weight value or a class
number.
– Weight display:
The current weight is shown in the
measured value line and the current
class in the text line.
– Display of classes:
The current class is shown in the
measured value line, and the current
weight in the text line.
Digital Input/Output Interface
The Classification application supports
the digital input/output-interface.
There are 4 control lines, or outputs,
which are activated as follows (see also
the diagram below):
– With 3 classes:
– Class 1
– Class 2
– Class 3
– Set
– With 5 classes:
– Classes 1/2
– Classes 2/3/4
– Classes 4/5
– Set
In Setup, under:
Appl:
W: 4.7.
Signum
®
3:
Appl 2
:
class: 4.7.
you can define whether these control
ports are
– switched off 
– always on 
– on at stability 
The “SET” output normally changes its
voltage level when the current weight
exceeds the minimum load.
Alternatively, you can assign the “Ready
for use” function to this port.
Configured in Setup under:
Appl:
W: 4.3.
Signum
®
3:
Appl 2
:
class: 4.3.
Digital I/O Interface
Control lines when working with 3 classes
Digital I/O Interface
Control lines when working with 5 classes
Class
1
23
Min. load
Max. load
Class 2
Set
Class 1
Class 3
Operative
Class
1
2345
Min. load
Max. load
Set
Classes 1/2
Classes 4/5
Operative
Classes 2/3/4