5. ELECTRICAL CONNECTIONS
WARNING
Prior to installing the X-ray system, it is advisable that all the electrical connections be in
place.
Timer electric connections
CAUTION
According to the relevant standard, the timer must be installed in a position allowing the
operator to always control the X-ray exposure.
On the timer installation wall, suitable conduits for the following electric cables must be
provided, pursuant to the electrical installation diagram (refer to §20 “ATTACHMENTS”):
a. Timer electrical cables
(refer to §5.9 “TIMER ELECTRIC CONNECTIONS”)
b. Cables connecting the timer and the tubehead
(refer to §5.9 “ TIMER ELECTRIC CONNECTIONS ”)
c. Cables connecting the timer and the RX signal lamp/s for external use:
X-Mind LIGHT (OPTIONAL)
d. Cables connecting the timer and the remote control button:
X-Mind ECB (IF PROVIDED)
Tubehead electrical connections
On the wall plate installation wall, a suitable conduit for the cable connecting the timer
and tubehead must be provided
(refer to §5.5 “CONNECTION TO THE POWER SUPPLY TERMINAL BOARD”)