17
- Reenabling the transmission depends on how the “rEACt” step has been set in the SET-
UP environment (passage by zero of the NET weight or instability of the NET weight of 20
divisions; by choosing “always” it works upon instability).
The data is transmitted using the standard string (StAb.St) or the extended string (StAb.EX); see
the “TRANSMISSION PROTOCOL” section for the description of the two strings
18.3 CONFIGURATION OF THE 2 SERIAL OUTPUT
PC SEL selection bidirectionnal output
CoM1 this RS485 (or RS232 output for model IPE50 Panel only, function of the position of the 3
jumpers on the PCB, RS485 in standard) is bidirectionnal (PC, PLC…), the output CoM2 RS232 serial
output is unidirectionnal (printer, repeater…)
CoM2 this RS232 output is bidirectionnal (PLC …), the output CoM1 (function of the position of
the 3 jumpers on the PCB, ) is unidirectionnal (printer, repeater …)
CoM.Prn unidirectionnal output
PrModE TRANSMISSION unidirectional (RS232 or 485 depend of the PC SEL choice)
Pr- no transmission disabled.
tPr enables the printing with ASCII printer (for example DP190 or TP200).
LP542P enables to print with labelling machine LP542PLUS
ALL.Std continuous transmission with standard string.
ALL.Eht continuous transmission with extended string.
PrPC.St transmission of the standard string upon the pressing of the ENTER key.
PrPC.EH transmission of the extended string upon the pressing of the ENTER key.
NOTE: if the “TOTALIZER” mode (horizontal or vertical) is active, the transmission through the
key is carried out by pressing the MODE key.
rEPE.6 transmission to 6-digit remote display IPE50.
(!) Pr-no
Nota: if the mode "totalization" is actived , the key MODE should be pressed.
bAud.Pr SET BAUD RATE
By pressing the ENTER key one accesses the selection of the data transmission speed
(measured in Baud = bit/second). The possible values are: 1200, 2400, 4800, 9600, 19200,
38400, 57600, 115200.
(!) 9600
bit.Pr SET PARITY, WORD, STOP BIT
By pressing the ENTER key one accesses the selection of the available values: n-8-1, n-8-2,
n-7-2, E-7-1, E-7-2.
(!) n-8-1
Prn.CtS RTS/CTS STATUS CONFIGURATION
On the printer serial line the indicator has a CTS input (Clear To Send). A device (like a
printer) that is slow in processing the data received, can interrupt the transmission temporarily
using this signal.
noCtS no signal
CtSL CTS active low (for LP542, TP200, DP24 printers)
CtSh CTS active high (for DP190 printers)
EmuCtS emulation of the CTS signal: one is asked to enter the number of characters
(nChrS), in 3 digits, which will be transmitted to the printer upon each