23
18.5 TRANSMISSION PROTOCOLS
The weight data transmission on the PC and PRT serial ports may take place in two formats:
STANDARD STRING
[CC]hh,kk,pppppppp,uu + CR + LF
in which: [CC] INSTRUMENT CODE IN THE FORMAT OF TWO ASCII DECIMAL DIGITS
ONLY IN THE CASE THAT THE 485 PROTOCOL IS SELECTED (FOR EXAMPLE 00).
hh UL Underload
OL Overload
ST Stability of the display
US Unstability of the display
TL Active inclination input
kk NT Net Weight
GS Gross Weight
GX Gross weight with sensitivity times 10
VL Value in microvolts relative to the weight
RZ Value in converter points relative to the weight
pppppppp 8 digits (including any sign and decimal point) which identify the weight. The
insignificant digits are filled with spaces. Through the MVOL and RAZF
command the indicator transmits the relative value on 10 digits instead of 8.
Uu Unit of measurement "kg" "bg" "bt" "lb" "mv" (microvolts) "vv" (converter points)
CR Carriage Return (13 ascii decimal character).
LF Line Feed (10 ascii decimal character).
The transmitted weight is the GROSS weight (GS) if no TARE WEIGHT has been entered; otherwise,
the NET WEIGHT (NT) will be transmitted.
EXTENDED STRING
(in response to the REXT command)
[CC]B,hh,NNNNNNNN,YYTTTTTTTT,PPPPPPPP,uu,(dd/mm/yybbhh:mm:ss"NO DATE TIME") +
CR + LF
in APW mode only 1 piece counting mode or in response to the REXT command)
[CC]B,hh,NNNNNNNN,YYTTTTTTTT,PPPPPPPP,AAAAA.AAAAA,uu + CR + LF
in which: [CC] INSTRUMENT CODE IN THE FORMAT OF TWO ASCII DECIMAL DIGITS
JUST IN CASE THE 485 PROTOCOL IS SELECTED (FOR EXAMPLE 00)
B scale number (1 in scale mode, 0 in pallet truck scale mode).