EasyManua.ls Logo

Scaime ipe 50 - Totalization, Counting, and Input;Output

Scaime ipe 50
30 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9
[II]ALRDXXXXX-YYYYYY <CR o CRLF>
In which: [II] 485 address (only when transmitting in 485 mode)
XXXXX rewriting number (from 00000 to 00255)
YYYYYY weigh number (from 000000 to 131072)
Instrument response:
[II]B,LLLLLLLLLLUU,YYTTTTTTTTTTUU<CR o CRLF>
In which: [II] 485 address (only when transmitting in 485 mode)
B scale number (always 1)
LLLLLLLLLL gross weight on 10 digits
UU unit of measure
YY spaces in the case of null or semiautomatic tare, PT in the case
of manual tare
TTTTTTTTTT tare weight on 10 digits
ALIBI MEMORY RESET (only with non approved instrument)
Command:
[II]ALDL <CR o CRLF>
In which [II] 485 address (only when transmitting in the 485 mode)
Instrument response:
[II]ALDLOK <CR o CRLF> if the cancellation has been effective
[II]ALDLNO <CR o CRLF> if the cancellation has not worked
The commands are ignored if the unit is not in the alibi memory functioning mode.
Nota: it could take several minutes to reset the data in the memory
17.4 INPUT/OUTPUT (in out)
Simple display functioning mode with in / out weighing function: the indicator acquires two weight values
through the confirmation of the operator and calculates the difference, automatically printing the data (if
the presence of a printer has been configured).
Once the in/out mode has been selected, the message “tyPE” is shown and one is asked to select with
ENTERthe printing mode of the acquired data:
- G.t. gross/tare:
GROSS Greater weight with unit of measure
TARE Lesser weight with unit of measure.
NET Difference between GROSS and TARE with unit of measure
- 1
st
.2
nd
first weigh/second weigh:
WEIGH 2 Second weight with unit of measure.
NET Difference without sign between WEIGH 1 and WEIGH 2 with unit of measure.
- in.out input/output:
INPUT First weight with unit of measure.
OUTPUT Second weight with unit of measure.
NET Zero weight with unit of measure >> if WEIGH 1 = WEIGH 2
INPUT NET >> if WEIGH 1 > WEIGH 2
Difference without sign between INPUT and OUTPUT with unit of measure.
OUTPUT NET >> if WEIGH 1 < WEIGH 2
Difference without sign between INPUT and OUTPUT with unit of measure.

Table of Contents