Print Command
[CC]PRNT<CR LF> or [CC]P <CR LF> (short command).
Instrument response: [CC]OK<CR LF> if the command has been RECEIVED; the instrument’s response
does not mean necessarily that the instrument executes the printout.
Command for viewing temporary message on the display
[CC]DISPNNVVVVVV <CR LF>
in which: NN: is the indicator display number, standard 00 (ascii hex)
VVVVVV is the message.
To cancel the message, send DISPNN or see DINT command
NOTES
In the case in which the display shown in the command is of the numeric type (for example the standard
display 00), if in the transmitted message there are two consecutive points the message is stopped after
the first of the two points. When the display is showing a message transmitted serially through the DISP
command, the indicator does not display those messages usually shown in the scale status (ZERO,
TARE, HOLD, ...).
Instrument response: [CC]OK<CR LF>
The message remains for the time set through the DINT command:
The ASCII characters having the decimal code greater than 31 are accepted.
With approved instrument:
One needs to wait the end of the current visualisation before being able to view the next one.
Command for setting display visualisation interval
[CC]DINTNNNN<CR LF>
in which: NNNN is the visualisation interval (in milliseconds), expressed in ascii hex character; for
example, in order to set a visualisation time of 2 seconds (2000 milliseconds, which converted into hex it
becomes 07D0), the command becomes DINT07D0<CR><LF>.
By setting a time equal to zero, the message transmitted with the DISP command remains permanently
shown on the display.
Instrument response: [CC]OK<CR LF>
With approved instrument:
The minimum settable time is 1 millisecond (0001HEX), and maximum settable time is 5 seconds (5000
milliseconds, 1388 HEX).
PC confirmation command
[CC]PCOK<CR LF>
The indicator shows on the display the "-PCOK-" message for about 2 seconds.
Instrument response: [CC]OK<CR LF>.
Serial command for setting the apw (only for the conuting operating mode)
[CC]SPMUvvvvvvvv <CR LF> or [CC]Xvvvvvvvv<CR LF> (short command).
in which: vvvvvvvv is the apw (up to 8 characters with decimal point); maximum value: 9999.999
System response: [CC]OK<CR LF>.
For example, to set a APW of 1.55 g, the command is the SPMU1.55<CR LF> or SPMU0001.550 <CR
LF> and all the various combinations adding zeros to the right or to the left but taking into consideration
that the maximum length of the APW field is 8 characters.
NOTES:
- The APW are not accepted in the SPMU.12<CRLF> format; these must be in the
SPMU0.12<CRLF> format.
- the APW are not accepted equal to zero.