8
- If the alibi memory is empty, when the MODE key is pressed the message “EMPTY” appears for
about a second, an error acoustic signal is enabled and one returns to weighing.
- If the entered ID is not valid, in other words, if there is no stored weigh relative to the entered ID, the
message " no id" appears and an error acoustic signal is enabled and one returns to weighing.
INITIALISATION OF THE ALIBI MEMORY
It is possible to cancel all the weighs made, initialising the alibi memory; this operation can be made
directly on the indicator (see the parameter “SETUP” >> “ini.AL” of the set-up environment, or through
the serial command (see “SERIAL COMMANDS” below).
NOTES:
- It is not possible to just cancel a single weigh.
The initialisation is possible only with a non approved instrument.
SERIAL COMMANDS
WEIGH STORAGE
Command
[II]PID<CRLF> or <ESC>[II]PID<STX>
in which: [II]: 485 address
<ESC>: 27 ascii decimal character
<STX>: 2 ascii decimal character
Instrument response to the [II]PID<CRLF> command:
[II]PIDSS,B,LLLLLLLLLLUU,YYTTTTTTTTTTUU,(ID | NO) <CRLF>
Instrument response to the <ESC>[II]PID<STX> command:
<ESC>[II]PIDSS,B,LLLLLLLLLLUU,YYTTTTTTTTTTUU,(ID | NO)<STX>
In which:[II] 485 address (only when transmitting in 485 mode)
SS OL" (weight in overload) or "UL" (weight in underload) or "ST" (stable weight) or
"US" (unstable weight) or "TL" (TILT input closed).
B scale number (always 1)
LLLLLLLLLL: gross weight on 10 digits
UU: unit of measure
YY: 2 spaces in the case of null tare or semiautomatic tare, “PT” in case of manual
tare
TTTTTTTTTT: tare on 10 digits
ID XXXX-YYYYYY in which: XXXXX is the rewriting number (5 digits, from
00000 to 00255) and YYYYYY is the weigh number (6
digits, from 000000 t0 131072).
In the case in which the gross weight is negative or unstable, the weight is transmitted but not the ID;
“NO” is in its place. In these cases there is no storage in the alibi memory.
Command to receive the date and time with the weight
[II]PIDD<CRLF> or <ESC>[II]PIDD<STX>
Instrument response to the [II]PIDD<CRLF> command:
[II]PIDSS,B,LLLLLLLLLLUU,YYTTTTTTTTTTUU,JJ/MM/AA HH:MM:SS (ID | NO) <CRLF>
date : JJ/MM/AA -> ex : 24/04/07 time : HH:MM:SS -> ex : 11:23:46
WEIGH READING
Command: