40 Schaeer Technologies
“The environment temperature is
above 70°C”
The ambient temperature is higher
than 70°C.
Switch o the device and wait
until the ambient temperature has
dropped below 70°C (158°F). If the
temperature is within the limit and
the error still occurs, please contact
your supplier.
“The mains current has exceeded
its limit, Attention: the ledge has
not been demagnetized!”
An eective current from the mains
is too high.
If the problem persists, contact
your supplier.
“The coil current has exceeded its
limit, Attention: the ledge has not
been demagnetized!”
An eective current through the coil
is too high.
Switch the device o and on and
try again.
If the problem persists, contact
your supplier.
“The capacitor current has
exceeded its limit, Attention: the
ledge has not been demagnetized!”
An eective current through the
capacitor is too high.
Switch the device o and on and
try again.
If the problem persists, contact
your supplier.
“The coil temperature is too low” The coil temperature -10°C. Switch o the device and wait until
the ambient temperature has risen
above -10°C (14°F).
If the temperature is within the limit
and the error still occurs, please
contact your supplier.
“The coil temperature is too high” The coil temperature 120°C. Switch o the device and wait
until the ambient temperature has
dropped below 120°C (248°F).
If the temperature is within the limit
and the error still occurs, please
contact your supplier.
“A coil current peak was detected,
Attention: the ledge has not been
demagnetized!”
A current peak has been detected. Switch o the device and wait a
few seconds before switching it on
again.
“A coil voltage peak was detected,
Attention: the ledge has not been
demagnetized!”
A voltage peak has been detected
exceeding 500V
Switch o the device and wait a
few seconds before switching it on
again.