EasyManua.ls Logo

Schaller RH2 - Adjustment behaviour of the sensor

Schaller RH2
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 23
5.3 Adjustment behaviour of the sensor
In humidity and temperature measurement, several parameters are responsible for the
adjustment behaviour (time until the actual measuring value is displayed). The para-
meter responsible for the highest measuring error is a temperature discrepancy
between the sensor resp. the whole measuring instrument and the material being
measured resp. the air.
Therefore, allow your humimeter device to adjust until the displayed temperature
corresponds to the actual temperature. The graph below shows how long it takes to
adjust from 20 °C to 30 °C.
To demonstrate the importance of temperature adjustment, the table below shows the
measuring errors due to a temperature difference between the measuring instrument
and the material being measured of only 1 °C / 1.8 °F at different ambient tempera-
tures.
10 °C (50 °F) 20 °C (68 °F) 30 °C (86 °F)
10 % r.h. +/- 0.7 % +/- 0.6 % +/- 0.6 %
50 % r.h. +/- 3.5 % +/- 3.2 % +/- 3.0 %
90 % r.h. +/- 6.3 % +/- 5.7 % +/- 5.4 %
At room temperature (20 °C / 68 °F) and an assumed humidity value of 50 % relative
humidity a temperature difference between the measuring sensor and the material
being measured of 1 °C / 1.8 °F causes a measurement error of 3.2 % relative humidi-
ty. A temperature difference of 3 °C / 5.4 °F would cause a measurement error of more
than 10 % relative humidity.

Table of Contents

Other manuals for Schaller RH2

Related product manuals