10.8.2 Configuring automatic switch-off
To do so: All of the options must be activated (see "10.5 Activating options").
1. Press
twice or hold for 2 seconds.
2. Select Options. To do so, press
or and confirm by pressing .
3. Select Auto Off Time. To do so, press
or and confirm by pressing .
4. Select the period of time you want the device to stay switched on (3 minutes,
5 minutes, 10 minutes, 20 minutes, 30 minutes). To do so, press
or and
confirm by pressing
.
» The setting has been saved.
5. Press
to leave the Options menu.
6. Press
to leave the main menu.
10.9 Calibrating the device
The calibration function is described in a separate operating manual.
10.10 Configuring the material calibration function
The type calibration function is described in a separate operating manual.
10.11 Online Send
To do so: All of the options must be activated (see "10.5 Activating options").
1. Press
twice or hold for 2 seconds.
2. Select Options. To do so, press
or and confirm by pressing .
3. Select Online Send. To do so, press
or and confirm by pressing .
» The setting has been saved.
» The device now automatically sends the stored measured value to the PC each
time the memory button is pressed.
4. Press
to leave the Options menu.
5. Press
to leave the main menu.