SERVOPRO NOx Analyzers Page
101
The Temperatures, Pressures and Electronic Pressure Controller (EPC) Voltage Percent
screens include the current device Values and the Low and High Alarm Limits. For
example, if the analyzer’s diode temperature drops lower than -5.50°C or exceeds -
4.50°C, an alarm will be triggered and displayed at the bottom of the
Measure Screen.
Pressures Screen
The Pressures screen displays current sample and air pressures and low and high alarm
limits in PSIG.
EPC Voltage Percent Screen
The EPC screen displays the percentage of EPC voltage being supplied to the EPC valve.