SERVOPRO NOx Analyzers Page
To access the Analyzer Digital Outputs status screen, press from the I/O Status
menu.
The Pin column indicates the connector and the pin number that are used to control the
digital output function. In the example (M7), M is the Main Connector on the analyzer’s
back panel, and 7 is the Pin Number on that connector.
The Status becomes closed when the function is true. In the example above, the analyzer
is in Range 4. Range 4’s status is closed. This will result in a closed contact to digital
ground. It can be measured from the Main Connector Pin 6 (digital ground) to the Main
Connector Pin 11.
ANALYZER DIGITAL INPUTS
→
→
→
→
NOTE: The Analyzer Digital Outputs screen allows the operator to view the status of
an analyzer output (Open or Closed) and where to find the corresponding pin number.
NOTE: These analyzer functions are not user-changeable and have static pin outs.
These screens are for viewing only.