SERVOPRO NOx Analyzers Page 16
5. If a channel does not exist, the answer for control and adjusting commands is, for
example, ATEM 0 3 NA. 3 is the number of the sub-channel.
6. If the device
is busy with a running function (for example, SLIN), every arriving control
command is ignored (except SRES and STBY), and the response message is e.g. SMAN
0 BS. If in the SINT mode an additional SINT KO command is received, the integrator is
reset to 0 and the integration is restarted.
7. If the command message contains data that the device cannot process (for example,
ESYZ K0 ABC), the response message is ESYZ 0 SE. A syntax error is recognized if the
data does not match the expected format or if the parameters do not fit the expected size.
8. Numbers are in floating-point format with decimal point. The decimal point can be
dropped for integers.
9. If you switch from Remote to Manual, the device
will remain in Manual mode until a
SREM K0 is received by the control computer. On the display, this mode is indicated by
REME (Remote Enable) on the Status line. In Manual mode, query commands via the
serial interface are possible at any time.
ABBREVIATIONS
Abbreviation Description
Mn Measuring range number (n)
Mn .. M4 Measuring range 1 .. 4
w.w .. ZZ. Numerical value
x Number
t Numeric integer value
a0 .. a4 Polynomial coefficients
s Status
yymmdd Date of format year, month and day with 2 characters each and no spaces
hhmmss Time of format hour, minute and second with 2 characters each and no spaces
In general, AK commands are subdivided into three classes: