SERVOPRO NOx Analyzers Page
172
x.x : NO
2
concentration
w.w : NO
x
concentration
Note: y.y, x.x, and w.w are normally 0.0 when
NO/NO
x
/NO
2
mode is selected.
t = Timestamp (1/10 sec.)
_AKON_K1 _AKON_s_z.z_t z.z : current measure value of channel 1.
Channel 1 is O
2
with option
_AKON_K4
_AKON_K5
_AKON_s_z.z_y.y_x.x_w.w_t Current measure value for all 4 ranges based on
their calibration offsets and gains.
t = Timestamp (1/10 sec.)
AMBE: Measuring Range Limit
Command Response Description
_AMBE_K0 _AEMB_s_M1_w.w
_M2_x.x
_M3_y.y
_M4_z.z
All existing measuring range limits.
AMBU: Upper and Lower Range Switchover Values for Auto Range
Command Response Description
_AMBU_K0 _AMBU_s_M1_w.w_W.W
_M2_x.x_X.X
_M3_y.y_Y.Y
_M4_z.z_Z.Z
Lower and upper range switchover value of auto
range.
APAR: Auto Calibration Tolerance Values
Command Response Description
_APAR_K0_SATK
_APAR_s_z.z_y.y_x.x_w.w
Auto calibration tolerance value (%):
z.z: Range 1
y.y: Range 2