EasyManua.ls Logo

Servomex SERVOPRO - Page 184

Servomex SERVOPRO
212 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SERVOPRO NOx Analyzers Page
183
y.y_M4_z.z
EKAK_K1_M1_
ww
_EKAK_s Sets the O
2
span gas values
EMBE: Set the Four Measuring Range Full scale Limits
Command Response Description
_EMBE_K0_M1_
w.w_M2_x.x_M3_
y.y_M4_z.z
_EMBE_s Sets the range full scale limits
EMBU: Set the Upper and Lower Range Switchover for Auto Range
Command Response Description
_EMBU_K0_M1_
w.w_W.W_M2_x.x_
X.X_M3_y.y_Y.Y_
M4_z.z_Z.Z
_EMBU_s Sets the lower and upper range switchover
limits
EPAR: Set Auto Calibration Tolerance Values
Command Response Description
_EPAR_K0_SATK_
z.z_y.y_x.x_w.w
_EPAR_s Auto calibration tolerance value (%):
z.z = Range 1
y.y = Range 2
x.x = Range 3
w.w = Range 4
ESYZ: Set System Time
Command Response Description
_ESYZ_K0_
yymmdd_hhmmss
_ESYZ_s Sets system time:
yymmdd: year, month, day

Related product manuals