SERVOPRO NOx Analyzers Page
hhmmss: hour, minutes, seconds
(each 2 characters, no spaces)
ET90: Set Lowpass Filter Time
Command Response Description
_ET90_K0_t
_ET90_s Sets lowpass filter time:
t = filter time
ETCP: Set TCP/IP Parameters
Command Response Description
_ETCP_K0_
zzz.zzz.zzz.zzz_
yyy.yyy.yyy.yyy_xxxx
_ETCP_s Sets TCP/IP parameters.
Zzz = TCP/IP address
yyy = TCP/IP subnet mask
xxxx = TCP/IP port
All changes take effect after next Power On
cycle.
EUDP: Set TCP/IP Data Streaming Parameters
Command Response Description
_EUDP_K0_<UDPPort>_
<DataFrequency>_
[<Mode>]_[<UDP_IP>]
_EUDP_s
Configures a UDP channel for data
streaming of the measuring values via
Ethernet UDP.
Port: port for opening the UDP connection.
Data Frequency: frequency for transmitting
the data in Hz.
Mode:
A: ASCII mode (optional).
UDP_IP: alternative IP address for opening
the UDP connection when it should not be