using the IP of the connected TCP/IP client
(optional).
Data Format
DATA is any number of AK commands delimited by a semicolon (;). Replace spaces in the AK
command with an underscore (_).
If data is given, UDP_IP has to be set to a legal IP address or a hyphen (-) if default access
should be used.
If data is omitted, “AKON K0” is used as default streaming data.
Format of the Streaming Data via UDP
ASCII Mode:
The measuring values will be sent with ASCII signs. The format is <sequence number> <data>
The sequence number will be incremented with every data packet that is sent.
<data> is the AK four-character code followed by the answer. See corresponding AK command
description.
Example
Sending “EUDP K0 7001 2 A – AKON_KO;ADUF_K0” will give the following streaming
result:
“123 AKON 4.07 901.33 22.50 3481639460 ADUF 4.30 4.59 4.45”, where 123 is the sequence
number.