pressures to be sure they are within their limits. If all diagnostic values are within their
limits, press the back button to return to the Manual Zero Calibration screen.
When the concentration value has stabilized, press to set the zero calibration. The
zero portion of the calibration should now be complete.
If the calibration was successful, the screen will say ***Saved Current*** above the
concentration value.
Example of a successful calibration.
If the calibration was unsuccessful, the screen will say Outside Deviation Limits.
Example of an unsuccessful calibration.