If the zero calibration is unsuccessful, check the following:
1. Make sure the correct gas was introduced into the analyzer.
2. Verify the Diagnostic Values while flowing gas.
3. Check Zero Gas Deviations under
Manual Deviations:
→
→
→
→
→
4. Check
Maximum Calibration Error under Calibration Setup:
→
→
→
→
→
After a successful Manual Zero Calibration, press the back button to return to the
Manual Calibration menu.