SERVOPRO NOx Analyzers Page
95
RANGE LIMITS
or
The operator can change the number of ranges and select a new specific full-scale
concentration in ppm. From the Range Setup menu press to access the Range Limits
screen.
To change the Range Limits from the factory settings, use the up or down arrows
to move the highlight to the field you intend to change. Press Enter to open
the field to change the value in ppm. Press Enter again to close the field. Press
to save your changes. To initiate the saved changes, press , then press and
select new ranges.
NOTES:
1. The Range Limit values must be set in ascending order.
NOTE: The analyzer is factory-configured with four physical ranges (Range 1 -Range
4).