SERVOPRO NOx Analyzers Page
96
2. The analyzer will not allow any of the range limits to exceed the maximum range
limit on the Range Limits screen. Example: Maximum Range Limit 3000.0.
3. To set a single range, set Range 1 to the desired value and all others to zero.
4. To set two ranges, set Range 1 to the lowest value, Range 2 to the highest value,
and the others to zero.
5. If new ranges are saved, the Auto Range Switch Points will be set to default
percentages of range limits. See
Auto Range Switch Points.