SERVOPRO NOx Analyzers Page
98
From the Range Setup menu press to access the Auto Range Switch Points screen.
The Default Switch Points are created by the range limits. The Up Switch Point is 90%
of the Range Limit. The Down Switch Point is 90% of the previous range’s Up Switch
Point.
To change the Auto Range Switch Points, use the up/down arrows to move
the highlight to the field you intend to change. Press Enter to open the field to
change the value in ppm. Press Enter again to close the field. Press to save
your changes. To initiate the saved changes, press , then press and select new
Auto Range Switch Points.
NOTE: Auto Range Switch Points determine when the analyzer automatically changes
a range up or down when the Auto Range function is turned on.