Chapter 2 Outline
Mobile DaRt Installation Manual
2-7
2.4 Conditions for Transportation and Storage
This condition is applied only at the time packed for transportation or
storage.
Transportation and Storage environment
Ambient temperature: -10℃~40℃
(The upper bound of the temperature is allowed up to +50℃ only for the
period of less than one month.)
Relative humidity: 10%~60%
Atmospheric pressure: 700hPa~1060hPa
When the unit is kept in storage for a long time until its installation, it is
recommended to charge the battery within the following period in order to
keep the performance of the battery:
Storage temperature 40℃: Once every 2 months
30℃: Once every 4 months
25℃: Once every 6 months
DO NOT CHARGE THE BATTERY WHEN PACKED.
Place the unit in a well-ventilated area to charge the battery.
NOTE
NOTE
CAUTION