8 OPERATION
8.5.6 Locking the display with a password
1. Log in to Exprt mode, see “8.2.2 Exprt mode”.
2. Activate or deactivate the display lock using CONFIG / LOCK.
If the display lock is activated:
The display is locked after a timeout of 5 minutes has elapsed
The display is unlocked temporarily by entering the password 000237
Note:
When the display is locked, only the congured measured value display can be seen.
8.5.7 Setting the medium adjustment
This setting is used to congure the speed of the medium adjustment. The medium
adjustment allows the sensor to adapt to changing measuring media. The slower this
adjustment occurs, the more stable the measurement.
Note:
The value is automatically set to 60 s when AUTCAL is performed.
Conguration
1. Log in to Exprt mode, see “8.2.2 Exprt mode”.
2. Set the parameter in the MEAS / ADAPT menu.
The possible values are 60 s, 30 s, 10 s, 5 s, 1 s.
If the medium remains the same, a value of 60 s is recommended.
8.5.8 Filtering measured values
Activating ltering
The lter supports the smoothing of the measured value; e.g., in the case of ripples on
level surfaces. For fast level changes, the average of the measured values over x sec-
onds is indicated.
1. Log in to Exprt mode, see “8.2.2 Exprt mode”.
2. Set the parameter in the MEAS / FILTER menu.
The possible values are Off, 500 ms, 1 s, 2 s, 5 s, 10 s.
8.5.9 Testing the conguration
Testing outputs
Switching and analog outputs can be simulated. This allows for the wiring and signal
values on the connected systems, such as a PLC, to be checked.