OPERATING INSTRUCTIONS | CFP CUBIC 8021989 / 2017-10-20 | SICK AG
Subject to change without notice
36
8 OPERATION
1
2
3
Fig. 6: CFP Cubic measuring range
1 Level
2 Offset
3 Measuring range
Note:
If the offset parameter is changed, the SPx/RPx/FLx/FHx/QaLow/QaHigh parameters
are automatically adjusted.
8.5.12 Resetting the calibration
Executing this function resets the empty adjustment that was carried out by the cus-
tomer, and the threshold value (EXPRT / MEAS / TRSHLD) and the lter time of the
medium adjustment (EXPRT / MEAS / ADAPT) are reset to the default value.
Conguration
1. Log in to Exprt mode, see “8.2.2 Exprt mode”.
2. Select MEAS / RESET and conrm using Set.
3. Conrm the RESET? safety prompt using Set.