OPERATING INSTRUCTIONS | CFP CUBIC8021989 / 2017-10-20 | SICK AG
Subject to change without notice
53
16FACTORY SETTING
16 Factory setting
Parameter Factory setting
SP1 80% of probe length measured from end of probe
RP1 5 mm below SP1
OUT1 Q1_Hno
SP2
For 5-pin versions: 20% of the probe length measured from the end of the probe
For 8-pin versions: 60% of the probe length measured from the end of the probe
RP2 5 mm below SP2
OUT2 Q2_Hno
TYP2 Q2_PNP
SP3 40% of probe length measured from end of probe
RP3 5 mm below SP3
OUT3 Q3_Hno
SP4 20% of probe length measured from end of probe
RP4 5 mm below SP4
OU4 Q4_Hno
TYP3 Q3_PNP
TYP4 Q4_PNP
QAHigh 50 mm below start of probe
QALOW 10 mm above end of probe
QAPOL QA_Nrm
QATYP Auto
QAFAIL 3.5 mA
SimCur SimOff
SimVol SimOff
DspVal Distan
Filter Off
SimLev SimOff
TrsHld 100
MaskZn 0 mm
MaskTr 50%
Mode Pulse
CalSta noCal
Probe/Type Depending on probe type: Rod/Rope
MaxCol Depending on measuring mode: HiSped = AnySped, HiAcc = 10 cm/s
MeasMd HiSpd
CalRng 6,005 mm
FomSta Inactive
Limit 90
Offset 0 mm
Unit mm
Lock Inactive