OPERATING INSTRUCTIONS | CFP CUBIC 8021989 / 2017-10-20 | SICK AG
Subject to change without notice
28
8 OPERATION
Conguration
Congure the Qx switching output as input (using Q1 as a redundant overll protection
example).
Note:
The switching output designation (Q1 in the example here) changes according to the
selected process variable to be output: Q1-LVL or Q1-TEMP is now Q1-INP.
1. Log in to Exprt mode, see “8.2.2 Exprt mode”.
2. Congure Q1 as input.
Set the parameter to INPUT in the Q1-### / MODE1 menu.
3. Congure the action to be performed when the signal is present.
Set the parameter to MAXLVL in the Q1-INP / ACTIO1 menu (if the external
signal is present, the CFP Cubic outputs the MAXLVL status).
4. Congure the form of the external signal.
Set the parameter to NO (normally open) in the Q1-INP / POL1 menu.
8.4 Conguration of analog outputs
8.4.1 Automatic signal detection
The CFP Cubic can automatically detect if a 4–20 mA or 0–10 V signal is required
using the connected output load.
The following rules apply:
4 mA to 20 mA if load < 500 ohms at U
V
> 15 V
4 mA to 20 mA if load < 350 ohms at U
V
> 12 V
0 V to 10 V if load > 750 ohms at U
V
≥ 14 V
Conguration
Note:
The analog output designation (QA in the example here) changes according to the
selected process variable to be output.
1. Perform automatic signal detection.
Set the parameter to AUTO in the QA-### menu.
Note:
Automatic signal detection is only active when the device is switched on for the rst
time. After this, the function can be activated again with AUTO in the QA-### menu.