11 TROUBLESHOOTING
Error Cause Solution
Sensor shows high level even though
the tank is empty.
The empty state of the probe has
changed signicantly due to altered
ambient conditions or aging
Increase the threshold so that faults are
below the threshold.
Check SigQu4, re-teach capacitance val-
ues in empty state, see “8.5.4 Perform-
ing empty adjustment”.
Deposit is detected as level Remove deposit.
If the level is detected at the deposit,
check SigQu1 and increase the thresh-
old until the level is no longer detected.
Strong temperature uctuations are
interfering with the probe
Increase the threshold, reduce the rate
of temperature change.