WARNING
Hazardous contact voltage
Risk of injury through hazardous contact voltage when the device is open or not completely
closed.
The degree of protection specied on the nameplate or in Technical data (Page59) is no
longer guaranteed if the device is open or not properly closed.
• Make sure that the device is securely closed.
The following safety note applies to the barrier box and the optics housings.
WARNING
Loss of explosion protection
Risk of explosion in hazardous areas if the device is open or not properly closed.
• Close the device as described in Installing/mounting (Page21).
The following safety notes apply to the optics housings.
WARNING
Hot surfaces
Risk of burns resulting from hot surfaces.
• Take corresponding protective measures, for example by wearing protective gloves.
WARNING
Alignment of optics housings in a hazardous area
Prior to operation align the optics housings as described in the detailed analyzer manuals. For
a correct alignment the Siemens sensor alignment kit must be used. The sensor alignment kit
is not certied for use in a potentially explosive atmosphere.
• When aligning the optics housings make sure that there is no potentially explosive
atmosphere (hot work permit)!
Commissioning
5.1General safety notes
LDS 6 & Optics housing CD 6 Compact Operating Instructions
48 Compact Operating Instructions, 12/2022, A5E03314905-AB