Climatic conditions
Ambient temperature
- during operation
- during transportation and storage
5...45°C (41...113°F)
-40...+70°C (-40...+158°F)
Atmospheric pressure 800...1100hPa (12...16psi)
Humidity <85% relative humidity, non-condensing
8.2 CD 6 optics housing for use in hazardous areas
LDS 6, Optics Housing CD 6 for use in
hazardous areas
Specication
Dimension, optics housing L: 450 mm (17 3/4"), Ø: 163mm (67/16"),
ange diameter DN80/PN16: 200 mm (7 7/8") or ANSI 4"/150
lbs.: 230 mm (9")
Dimension, purging tube L: 400, 800 or 1 200 mm (15 3/4, 31 1/2 or 47 1/4"), Ø: 44 mm
(13/4")
Weight 2 x 11kg (24lbs.)
Power supply According to the intrinsically-safe standards (IEC/
EN60079-11)
Power consumption Max. 0.6W
Degree of protection IP65 according to EN60529
EMC - Electromagnetic compatibility According to EN61326 and standard classication of NA‐
MURNE21
Electrical safety According to EN61010-1, overvoltage classication II.
Permissible ambient temperature (oper‐
ation) for the components of the Ex-rel‐
evant IParea
-20...+60°C (-4...+140°F)
Permissible ambient temperature (stor‐
age and transport)
-30...+70°C (-22...+158°F)
Atmospheric pressure 800...1100 hPa (12...16psi)
Maximum allowable work pressure
(MAWP)
For pressure-resistant window ange 1.6MPa (232psi) at
20°C (68°F)
For standard ange 1050hPa (15psi) at 20°C (68°F)
Permissible ambient humidity <95% relative humidity, non-condensing
Optics housing/process interface DN65/PN6 and ANSI4"/150 lbs.;
DN80/PN16 (only as pressure-resistant variant)
Measurement path length 0.3...12m (1...39ft)depending on measuring conditions
(dust load)
Dust load Depending on particle size distribution and optical path length
See also
Safe installation (Page24)
Technical data
8.2CD 6 optics housing for use in hazardous areas
LDS 6 & Optics housing CD 6 Compact Operating Instructions