Opdimar Register 3 SPB7-230.033.12 Page 4 of 26 Siemens-Elema AB
Installation Rev. 01 09.02 SPS-UD Solna, Sweden
5 - 4 Installation
Installation of biopsy controller holders 5
The biopsy controller is attached to the rear cover of the stand. To install the holders
proceed as follows:
1. Attach the two drilling templates (1/Fig. 4) included in this document on the inside
of the rear cover according to Fig. 4. The template marked B shall be placed
underneath the template marked A with no space between them. Make sure that
the templates are centered and that the sides of the templates are parallel to the
rear cover sides.
2. Drill five holes with a 5 mm drill in the rear cover according to the templates.
3. Install the three holders (2/Fig. 4) by using the nuts (3/Fig. 4).
4. Install the two standoffs (4/Fig. 4) by using the screws (5/Fig. 4) and the washers
(6/Fig. 4).
UP
OPD00267
985 – 5 mm
3
4
5
2
1
A
B
6
Fig. 4 Rear cover, from the back side