Prerequisites 1 - 3
Siemens-Elema AB Register 3 SPB7-230.033.12 Page 3 of 4 Opdimar
Solna, Sweden Rev. 01 09.02 SPS-UD Installation
Documents required 1
• Supplement to the Instructions for Use MAMMOMAT 3000 - Opdimar
(included in the Opdimar delivery)
• MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
(included in the Opdimar delivery)
• MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy
Attachment and Opdimar DHHS Maintenance Instructions
(included in the Opdimar delivery)
• MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy
Attachment and Opdimar Measurement Certificates
(included in the Opdimar delivery)
• MAMMOMAT 1000/3000/3000 Nova - Opdimar Planning Guide
• Installation Instructions for Isolation P.c. Board D707, RX B7-120.031.04. .. (where
applicable)
• MAMMOMAT 1000/3000/3000 Nova Installation and Start-Up Instructions
• MAMMOMAT 1000/3000 Nova Wiring Diagram
• MAMMOMAT 3000 Installation and Setting Instructions Update Kit for Stereotactic
Biopsy Attachment (where applicable)
CD-ROMs required 1
• Ultra 10 Service Manual, part No. 65 27 670 (included in the Opdimar delivery)
Meters and appliances required 1
• Protective ground wire tester (44 15 899 RV090)
• Service PC (e.g. Siemens-Nixdorf PCD 3NSX/20 or similar) with connecting cable, PC to
generator (part No. 99 00 440 RE999)
• Stereo calibration phantom (included in the delivery) (part No.64 30 701)
• AEC calibration plexiglass, four plates measuring 150 mm x 150 mm x 19 mm and one
plate measuring 150 mm x 150 mm x 9,7 mm, part No. 65 61 232 and 65 61 224
respectively
• Resolution phantom with at least 10 line pairs per mm
• 4.5 cm PMMA
Tools required 1
• Standard installation tools
• Electric drill with 5 mm drill
Additional installation material required 1
• Extension mains cord with at least three outlets, of protective earth type