Installation 5 - 13
Siemens-Elema AB Register 3 SPB7-230.033.12 Page 13 of 26 Opdimar
Solna, Sweden Rev. 01 09.02 SPS-UD Installation
Cables to biopsy unit 5
The cable with connectors marked X207 and X885 is laid between the biopsy controller
and the biopsy unit connector of the MAMMOMAT stand.
1. Connect the cable to X207 on the biopsy controller and X885 on the stand, see
Fig. 12 (serial number higher than 3000) or Fig. 13 (serial number lower than
3000).
Power connection 5
The mains cable supplies the biopsy controller with 230 V AC from the stand. Together
with the mains cable a protective earth cable shall be installed.
The connection between the biopsy controller and the stand can be made in two ways
depending on MAMMOMAT serial number:
Serial number higher than 3000: biopsy controller mains connector to MAMMOMAT
3000 connector Z801, see Fig. 12
Serial number lower than 3000: biopsy controller mains connector to MAMMOMAT
connector X881, see Fig. 13
1. Connect the cables, mains cable and protective earth cable, to Z801 or X881 on
the stand.
The brown cord (1/Fig. 12) to Z801-1, the blue cord (2/Fig. 12) to Z801-2 and the
green-yellow cord (3/Fig. 12)to X880-4 for MAMMOMAT serial number higher
than 3000. For MAMMOMAT serial number lower than 3000, connect according to
Fig. 13. Strain relieve the cable with the supplied strain relief clamp.
2. Connect the three-pole connector of the mains cable to the biopsy controller, see
Fig. 11.
The evaluation unit can not be used with Opdima.
If an evaluation unit is connected to the MAMMOMAT this must be
disconnected at X885.
The evaluation unit can not be used with Opdima.
If an evaluation unit is connected to the MAMMOMAT, the power
connection of the evaluation unit must be disconnected at the
stand.
NOTICE
NOTICE
OPD00474
1 (BN)
2 (BU)
3 (GNYE)
X885
Z801
Protective earth cable
Fig. 12 Connection of mains cable, protective earth cable and X885, serial number higher than 3000