EasyManua.ls Logo

Siemens MAMMOMAT 1000 - Restoring Installation Parameters

Siemens MAMMOMAT 1000
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Opdimar Register 3 SPB7-230.033.12 Page 8 of 26 Siemens-Elema AB
Installation Rev. 01 09.02 SPS-UD Solna, Sweden
5 - 8 Installation
Restoring installation parameters 5
If the software has been changed, the installation parameters have to be restored.
Proceed as follows:
1. Switch on the MAMMOMAT. Install the service PC, (see Working with the Service
PC in MAMMOMAT 1000/3000/3000 Nova Installation and Start-Up Instruction).
2. Start the service program with the new service diskette.
3. Select
Main menu => Backup => Copy disk to installation area => All
. This will
copy the previously stored data from the hard disk to the installation area of the
MAMMOMAT.
4. Select
Main menu => Configuration => Power
and note the values for all four
focus.
5. Select
Main menu => Configuration => Anode
and enable/disable the W
(tungsten) anode depending on the system configuration.
6. Select
Main menu => Configuration => Power
. Make sure that the values are
identical to the values noted in step 3. above, if not correct them.
7. Select
Main menu
=>
Configuration => Mischelaneous=> Cassette loaded switch
and set the switch to ON.
8. Switch off the MAMMOMAT with the main circuit breaker.
For serial numbers lower than 5000 the W anode should always
be enabled.
For serial numbers higher than 5000 it depends on what the cus-
tomer ordered, if the W anode should be enabled or disabled. This
can be checked by looking at the filter disk in the collimator. If the
filter disk has only two filters and a lead strip mounted instead of
a third filter, the W anode should be disabled.
NOTE
NOTE

Table of Contents

Other manuals for Siemens MAMMOMAT 1000

Related product manuals