Opdimar Register 3 SPB7-230.033.12 Page 8 of 26 Siemens-Elema AB
Installation Rev. 01 09.02 SPS-UD Solna, Sweden
5 - 8 Installation
Restoring installation parameters 5
If the software has been changed, the installation parameters have to be restored.
Proceed as follows:
1. Switch on the MAMMOMAT. Install the service PC, (see Working with the Service
PC in MAMMOMAT 1000/3000/3000 Nova Installation and Start-Up Instruction).
2. Start the service program with the new service diskette.
3. Select
Main menu => Backup => Copy disk to installation area => All
. This will
copy the previously stored data from the hard disk to the installation area of the
MAMMOMAT.
4. Select
Main menu => Configuration => Power
and note the values for all four
focus.
5. Select
Main menu => Configuration => Anode
and enable/disable the W
(tungsten) anode depending on the system configuration.
6. Select
Main menu => Configuration => Power
. Make sure that the values are
identical to the values noted in step 3. above, if not correct them.
7. Select
Main menu
=>
Configuration => Mischelaneous=> Cassette loaded switch
and set the switch to ON.
8. Switch off the MAMMOMAT with the main circuit breaker.
For serial numbers lower than 5000 the W anode should always
be enabled.
For serial numbers higher than 5000 it depends on what the cus-
tomer ordered, if the W anode should be enabled or disabled. This
can be checked by looking at the filter disk in the collimator. If the
filter disk has only two filters and a lead strip mounted instead of
a third filter, the W anode should be disabled.
NOTE
NOTE