6 - 1
Siemens-Elema AB Register 3 SPB7-230.033.12 Page 1 of 4 Opdimar
Solna, Sweden Rev. 01 09.02 SPS-UD Installation
Final procedures 6
General 6
The following procedures shall be performed:
• Starting up the workstation, see Page 6 - 2
• Country settings,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions
• Test of units,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions
• CCD camera calibration,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions
• Calibration of the biopsy unit,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions
• Verifying the calibration of the biopsy unit,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar
Service Instructions
• Check of Opdimar AEC,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar
Service Instructions
• Check of resolution,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar
Service Instructions
• Activate DICOM (optional),
see DICOM option in chapter Service mode in MAMMOMAT 1000/3000/3000 Nova -
Opdimar Service Instructions
• Network settings,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions
• Printer setup (optional),
see Use of printer setup in chapter Service mode in MAMMOMAT 1000/3000/3000 Nova
- Opdimar Service Instructions
• Mounting of covers, see Page 6 - 2
• Protective earth measurement,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar
Service Instructions
• Fill in the IVK list or report the IVK components to your responsible uptime service center
(USC).
• Performing of DHHS maintenance, see Page 6 - 3
The Opdimar external diaphragm must be used for tests and cali-
brations involving radiation.
NOTICE