Opdimar Register 3 SPB7-230.033.12 Page 16 of 26 Siemens-Elema AB
Installation Rev. 01 09.02 SPS-UD Solna, Sweden
5 - 16 Installation
Installation of cable duct 5
Four cables are laid from the bottom of the biopsy controller in a cable duct to the cable
outlet cover (3/Fig. 17). To install the cable duct on the rear side of the stand proceed as
follows:
1. Attach the cable duct (2/Fig. 17) to the rear cover by removing the protective strips
from the attached Velcro strips and place the duct flush with the biopsy controller
(1/Fig. 17). Secure the cable duct to the two standoffs (5/Fig. 17) with the
delivered screws and washers.
Modification of cable outlet cover 5
To allow the cables to pass through the cable outlet cover, the cover has to be modified
according to the following:
1. Cut an opening centrally on the cable outlet cover (4/Fig. 17). The opening should
be approximately 50 x 25 mm.
2. Attach the supplied plastic strip around the opening of the cable outlet cover
(4/Fig. 17).
3. Install the cable outlet cover (4/Fig. 17).
4. Mount the previously removed stand covers.
UP
OPD00164
4
3
1
2
50m
m
25mm
5
Fig. 17 Cable duct