Installation 5 - 7
Siemens-Elema AB Register 3 SPB7-230.033.12 Page 7 of 26 Opdimar
Solna, Sweden Rev. 01 09.02 SPS-UD Installation
Replacing software 5
MAMMOMAT with serial number lower than 5200 must be updated with new software.
1. Replace the existing software on boards D740, D701 and D702 (if not already
done in previous steps) with the V2.2 (or higher) software.
Installing isolation board D707 5
If not already done in conjunction with installation of printer (option), an isolation board
D707 (see Fig. 10) shall be installed. Install the board according to Installation Instructions
for Isolation P.c. Board D707, RX B7-120.031.03. ...
If upgrading from a version lower than V1.5, the AEC have to be
upgraded, see Update Instruction, UI 001/96.
OPD00154
D702
Fig. 7 Board D702