Opdimar Register 3 SPB7-230.033.12 Page 10 of 26 Siemens-Elema AB
Installation Rev. 01 09.02 SPS-UD Solna, Sweden
5 - 10 Installation
Cables to MAMMOMAT generator (and printer if present) 5
To lay the cables, proceed as follows:
1. Disconnect the printer cable (if connected) from the MAMMOMAT generator.
2. Open the cable duct between the generator and the stand and lift up the cable (if
present).
The cable will be reused to connect the printer to X205 on the biopsy controller.
3. Place the new cable with connectors marked X206 and X2 in the cable duct
between the generator and the biopsy controller. Use compartment according to
the figure below.
OPD00156
Cable X206-X2
Serial number higher than 3000
Serial number lower than 3000
Fig. 9 Cable-duct compartments